Suzhou Hangruisi Measurement and Control Technology Co., Ltd.

Force Control System Solutions Provider

+
  • 109-1.jpg

HRS-D505 Industrial Control Instrument

Category:

Instrumentation/Transmitter Series

Display instrument

  • Product Description
  • Download Data

    Product features

    ● Used in applications that require converting force measurement signals into digital communication signals, standard analog output signals, and simple control.

    ● Analog output signals can be selected from 0–20 mA, 4–20 mA, 0–5 V, or 1–5 V; the digital communication interface can be RS‑485 or RS‑232, both with isolation.

    ● The controller features upper‑limit, mid‑range, and lower‑limit threshold detection; for detailed control functions, please refer to Note 3 in the user manual.

    ● The limit output offers three operating modes: upper‑limit, lower‑limit, and both‑upper‑and‑lower‑limit detection. For detailed explanations, please refer to Note 2 in the user manual.

    ● The controller is powered by a 24 V DC supply with a voltage tolerance of ±5 V, ensuring greater safety and stability.

     

    Technical specifications

    *1. Display window: dual-row 6-digit LED display, with character heights of 9 mm and 7.5 mm, respectively.

    *2. Graduation values: 1, 2, 5, 10, 20, 50

    *3. Maximum weighing display range: 999,999

    *4. Decimal place: 0, 0.0, 0.00, 0.000, 0.0000, 0.00000

    *5. Static accuracy class: Resolution 900,000, accuracy 0.01%

    *6. Maximum signal input range: -3.6 mV/V to 3.6 mV/V (equivalent to -18 mV/V to 18 mV/V)

    *7. Zero Drift: ≤0.05 μV (@0.02 mV/V)

    *8. Span temperature coefficient: ≤10 ppm/°C

    *9. Sensor interface input impedance: ≥20 MΩ

    *10. Nonlinearity error: ≤0.002%FS

    *11. A/D conversion rate: 400 samples per second

    *12. Zero Drift: ≤10 μV/°C

    *13. Span Temperature Coefficient: ≤0.02%FS/°C

    *14. Sensor Type: Strain-Gauge Sensor

    *15. Sensor excitation voltage: DC 5 V, 100 mA; up to four 350 Ω sensors can be connected in parallel.

    *16. Digital output (contact) capacity: Relay output capacity: AC 220 V, 1 A, 6 channels; Transistor output capacity: DC 24 V, 0.5 A.

    *17. Digital input voltage: DC 24 V

    *18. Digital input current: 4–6 mA

    *19. Power supply range: DC 24 V (±5)

    *20 Product power: ≤10W

    *21. Operating temperature: -10℃ to 50℃

    *22. Temperature range: ≤90% relative humidity (no condensation)

     

    Load direction

     